Difference between revisions of "Bioflo 110 Fermentor - New Brunswick Scientific"
From StemNode Lit Reviews
(Created page with "---------- <!-- This shows the authors that have edited a page and gives them credit for their contribution --> {{Template:CiteAuthors}} '''This system is also called:''' '''...") |
|||
| Line 3: | Line 3: | ||
{{Template:CiteAuthors}} | {{Template:CiteAuthors}} | ||
| − | '''This system is also called:''' | + | '''This system is also called: Bioreactor''' |
| − | '''Keywords:''' | + | |
| + | '''Keywords: Fermentation, Anaerobic, Aerobic''' | ||
== Operation == | == Operation == | ||
| − | + | Control can be performed from either the primary control unit, or an attached PC | |
| − | |||
===Troubleshooting === | ===Troubleshooting === | ||
| Line 20: | Line 20: | ||
! Problem | ! Problem | ||
! Symptom (s) | ! Symptom (s) | ||
| − | ! Fundamental Cause | + | ! Fundamental Cause (s) |
| − | ! Solution | + | ! Solution(s) |
|- | |- | ||
| − | | | + | | Drifting pH probe |
| − | | | + | | Hard to get a steady reading upon calibration |
| − | | | + | | Air bubbles may prevent contact between the the half cell and probe fluid |
| − | | | + | | Check for bubbles at the tip of the probe. Carefully swing in circles with tip of probe outward to force bubbles toward the top of the probe. Alternatively, probe may need replacement. |
| + | |- | ||
| + | |Primary Control Unit not starting up fully | ||
| + | |Partial or fully dead screen. Failure to fully boot up the PCU. | ||
| + | |Weak connection at the power supply due to corrosion and possibly overheating if ran for long periods. | ||
| + | |Check all connections in the PCU, especially to and from the power supply, for blackened/burnt wires or corrosion. Grind away anything that may impede a good connection. | ||
| + | |- | ||
| + | |Quick draining of connected air tanks | ||
| + | |Air tanks drain much faster than expected by calculation | ||
| + | |Leak in Mass flow controller. | ||
| + | |Open up the mass flow controller and identify the point of leakage. Spraying with soap is an option but care for electrical connections should be taken. | ||
| + | |- | ||
| + | |Improper pH control | ||
| + | |The pump module is operating but pH is not adeqautely adjusting | ||
| + | |A clog in the buffer line. This is common for thin tubing where salting out of any pH adjustment solution could occur. | ||
| + | |Try and break leak with tubing, cut leak out of tube and rejoin tubing together, or replace tubing. | ||
|} | |} | ||
| + | |||
| + | ==== Operating Procedures ==== | ||
==Maintenance == | ==Maintenance == | ||
| Line 37: | Line 54: | ||
{| class="wikitable sortable" | {| class="wikitable sortable" | ||
|- | |- | ||
| + | !Action | ||
! Frequency | ! Frequency | ||
| − | |||
! Who Performs? | ! Who Performs? | ||
! Time to Complete | ! Time to Complete | ||
|- | |- | ||
| − | | | + | |Replace pH probe storage solution |
| − | | | + | | Monthly |
| − | | | + | | User |
| − | | | + | | 5 min |
|} | |} | ||
==Repair == | ==Repair == | ||
| + | |||
| + | == Accessories == | ||
== Safety ==<!--Personal protective equipment, or operating requirements --> | == Safety ==<!--Personal protective equipment, or operating requirements --> | ||
Revision as of 04:47, 5 March 2019
Authors:
This system is also called: Bioreactor
Keywords: Fermentation, Anaerobic, Aerobic
Contents
Operation
Control can be performed from either the primary control unit, or an attached PC
Troubleshooting
| Problem | Symptom (s) | Fundamental Cause (s) | Solution(s) |
|---|---|---|---|
| Drifting pH probe | Hard to get a steady reading upon calibration | Air bubbles may prevent contact between the the half cell and probe fluid | Check for bubbles at the tip of the probe. Carefully swing in circles with tip of probe outward to force bubbles toward the top of the probe. Alternatively, probe may need replacement. |
| Primary Control Unit not starting up fully | Partial or fully dead screen. Failure to fully boot up the PCU. | Weak connection at the power supply due to corrosion and possibly overheating if ran for long periods. | Check all connections in the PCU, especially to and from the power supply, for blackened/burnt wires or corrosion. Grind away anything that may impede a good connection. |
| Quick draining of connected air tanks | Air tanks drain much faster than expected by calculation | Leak in Mass flow controller. | Open up the mass flow controller and identify the point of leakage. Spraying with soap is an option but care for electrical connections should be taken. |
| Improper pH control | The pump module is operating but pH is not adeqautely adjusting | A clog in the buffer line. This is common for thin tubing where salting out of any pH adjustment solution could occur. | Try and break leak with tubing, cut leak out of tube and rejoin tubing together, or replace tubing. |
Operating Procedures
Maintenance
Maintenance Schedule
| Action | Frequency | Who Performs? | Time to Complete |
|---|---|---|---|
| Replace pH probe storage solution | Monthly | User | 5 min |
Repair
Accessories
Safety
Versions of the System
| Version | Dates of Production | End of Production | Number Produced |
|---|---|---|---|
| (Insert Version) | (State a Date) | (State a Date) | (State a unitless quantity) |
'