Bioflo 110 Fermentor - New Brunswick Scientific

From StemNode Lit Reviews

Authors:

This system is also called: Bioreactor

Keywords: Fermentation, Anaerobic, Aerobic


Operation

An overview of operations is given here; https://www.che.utah.edu/undergraduate/projects_lab/modular_benchtop_fermentor/

Troubleshooting

Problem Symptom (s) Fundamental Cause (s) Solution(s)
Drifting pH probe Hard to get a steady reading upon calibration Air bubbles may prevent contact between the the half cell and probe fluid Check for bubbles at the tip of the probe. Carefully swing in circles with tip of probe outward to force bubbles toward the top of the probe. Alternatively, probe may need replacement.
Primary Control Unit not starting up fully Partial or fully dead screen. Failure to fully boot up the PCU. Weak connection at the power supply due to corrosion and possibly overheating if ran for long periods. Check all connections in the PCU, especially to and from the power supply, for blackened/burnt wires or corrosion. Grind away anything that may impede a good connection.
Quick draining of connected gas tanks Gas tanks drain much faster than expected by calculation. Hissing in the mass flow controller is louder than expected. Leak in Mass flow controller. Open up the mass flow controller and identify the point of leakage. Spraying with soap is an option but care for electrical connections should be taken.
Improper pH control The pump module is operating but pH is not adeqautely adjusting A clog in the buffer line. This is common for thin tubing where salting out of any pH adjustment solution could occur. Try and break leak in the tubing, cut leak out of tube and rejoin tubing together, or replace tubing.
Failure to pump Pump heads are rotating but fluid is not moving Pump head may be out of alignment due to a breakage on the inside. This keeps the pump head from being able to exert pressure on the tubing. Open the pump moduel and identify broken pieces. Super glue or replace broken parts.

Operating Procedures

Maintenance

Maintenance Schedule

Action Frequency Who Performs? Time to Complete
Replace pH probe storage solution Monthly User 5 min

Repair

Accessories

Safety

Versions of the System

Version Dates of Production End of Production Number Produced
(Insert Version) (State a Date) (State a Date) (State a unitless quantity)

Upgrades, Modifications, Hacks

Diagrams

Warranties

Certifications and Training

Manufacturer Information